CFD ANALYSIS OF MINI VENTURI-TYPE AIR FLOWMETERS FOR LAB-SCALE APPLICATIONS

Being one of the cheapest devices for air flow measurement within a wide range of flowrates, Venturi-type air flowmeters (VTAFs) can also be used for measuring very small air flowrates, usually less than 50 to 100 liters per minute, which is usually encountered in lab-scale applications. This study presents a model-based evaluation of VTAFs for lab-scale applications. The pressure and velocity fields in mini VTAFs were calculated by computational fluid dynamics (CFD) model using Ansys Fluent™. Results showed that sensitive and accurate measurements of volumetric flowrate can be achieved using mini VTAFs. A diameter ratio of 0.6 for the venturi throat provides a minimum detection limit of 2.46 LPM while an overall pressure drop of around 115 Pa at an air flowrate of 50 LPM at 20°C.

___

  • [1] Kumar P., San S.M., “CFD Study of the Effect of Venturi Convergent and Divergent Angles on Low Pressure Wet Gas Metering”, J. Appl. Sci., 14, 3036-3045, 2014.
  • [2] Zhang F., Dong F., Tan C., “High GVF and low pressure gas-liquid two-phase flow measurement based on dual-cone flowmeter”, Flow Meas. Instrum. 21, 410-417, 2010.
  • [3] Demir S., “A Practical Model for Estimating Pressure Drop in Cyclone Separators: An Experimental Study”, Powder Technol. 268, 329-338, 2014.
  • [4] Demir S., Karadeniz A., Aksel, M., “Effects of cylindrical and conical heights on pressure and velocity fields in cyclones”, Powder Technol., 295, 209-217, 2016
  • [5] Demir S., Karadeniz A., Manav Demir N., Duman S., “Excel VBA-Based Solution to Pipe Flow Measurement Problem”, eJSIE, 10(3), 1, 2018.
  • [6] Huang X., van Sciver S.W., “Performance of a Venturi Flow Meter in Two-Phase Helium Flow”, Cryogenics, 36, 303-309, 1996.
  • [7] Steven R., “Wet Gas Metering with a Horizontally Mounted Venturi Meter”, Flow Meas. Instrum., 12, 361-372, 2002.
  • [8] Lide F., Tao Z., Ningde J., 2007, “A Comparison of Correlations Used for Venturi Wet Gas Metering in Oil and Gas Industry”, Journal of Petroleum Science and Engineering, 57, 247-256, 2007.
  • [9] Steven R., “A Dimensional Analysis of Two Phase Flow Through a Horizontall Installed Venturi Flow Meter”, Flow Meas. Instrum., 19, 342-349, 2008.
  • [10] Steven R., Hall A., “Orifica Plate Meter Wet Gas Flow Performance”, Flow Meas. Instrum., 20, 141-151, 2009.
  • [11] Kumar P., Bing M.W.M., “A CFD Study of Low Pressure Wet Gas Metering Using Slotted Orifice Meters”, Flow Meas. Instrum., 22, 33-42, 2011.
  • [12] He D., Bai B., “A New Correlation for Wet Gas Flow Rate Measurement with Venturi Meter Based on Two-Phase Mass Flow Coefficient”, Measurement, 58, 61-67, 2014.
  • [13] Monni G., de Salve M., Panella B., “Two_phase Flow Measurements at High Void Fraction by a Venturi Meter”, Progress in Nuclear Energy 77, 167-175, 2014.
  • [14] Gajan P., Decaudin Q., Couput J.P., “Analysis of High Pressure Tests on Wet Gas Flow Meterin with a Venturi Meter”, Flow Meas. Instrum. 44, 126-131, 2015.
  • [15] Xu Y., Zhang Q., Zhang T., Ba X., “An Overreading Model for Nonstandard Venturi Meters Based on H Correction Factor”, Measurement, 61, 100-106, 2015.
  • [16] Pan Y., Hong Y., Sun Q., Zheng Z., Wang D., Niu P., “A New Correlation of Wet Gas Flow for Low Pressure with a Vertically Mounted Venturi Meter”, Flow Meas. Instrum., 70, 101636, 2019.