Objective: The aim of this study was to evaluate the performance of dual-mobility (DM) cup systems for revision total hip arthroplasty (rTHA) in patients who had high risk for instability. Methods: We prospectively followed up 34 hips of 30 patients (27 females, 3 males; mean age: 66.1 (range: 33 to 89) years) who underwent rTHA with DM cups for aseptic loosening in 23 hips, infection treatment as second or single stage in nine hips, and instability in two hips. Clinical functions of the patients were evaluated using the Harris Hip Score (HHS), and radiological migration or loosening of the DM cups were recorded. The survival of the components was calculated with the Kaplan-Meier survival analysis and failure was defined as any dislocation of the polyethylene (PE) insert, intraprosthetic dislocation (IPD), aseptic loosening of any component or total hip system revision due to any reason. Results: The mean duration of follow-up was 3.52 (range: 2.05 to 6.26) years. There was one dislocation of PE insert (2.9%), which was treated with closed reduction. There were two (5.8%) re-revisions for cemented DM cup due to migration. There was one PE insert and head change due to subacute infection. The mean HHS increased from 42.8 ± 6.7 (range: 34 to 60) points preoperatively to 87.3 ± 5.8 (range: 75 to 98) points postoperatively. The cumulative survival rate of the DM cup system was 91.2% (95% CI: 81.6 e100%) with any revision, 94.1% (95% CI: 86.2e100%) with aseptic loosening and %97.1 (%95 CI: 91.4 e100%) with dislocation as the end point at 3.5 years. Conclusion: Dual-mobility cups may provide good stability and represent a good option for revision acetabular reconstruction in patients who have high risk for instability.
___
1. Springer BD, Fehring TK, Griffin WL, Odum SM, Masonis JL. Why revision total hip arthroplasty fails. Clin Orthop Relat Res. 2009;467:166e173.
2. Plummer DR, Christy JM, Sporer SM, Paprosky WG, Della Valle CJ. Dual-mobility articulations for patients at high risk for dislocation. J Arthroplasty. 2016;31: 131e135.
3. Parvizi J, Picinic E, Sharkey PF. Revision total hip arthroplasty for instability: surgical techniques and principles. J Bone Joint Surg Am. 2008;90:1134e1142.
4. Berend KR, Lombardi Jr AV, Mallory TH, Adams JB, Russell JH, Groseth KL. The long-term outcome of 755 consecutive constrained acetabular components in total hip arthroplasty. Examining the successes and failures. J Arthroplasty. 2005;20:93e102.
5. Kung PL, Ries MD. Effect of femoral head size and abductors on dislocation after revision THA. Clin Orthop Relat Res. 2007;465:170e174.
6. Saadat E, Diekmann G, Takemoto S, Ries MD. Is an algorithmic approach to the treatment of recurrent dislocation after THA effective? Clin Orthop Relat Res. 2012;470:482e489.
7. Plummer DR, Haughom BD, Della Valle CJ. Dual mobility in total hip arthroplasty. Orthop Clin North Am. 2014;45:1e8.
8. Noyer D, Caton JH. Once upon a time.... Dual mobility: history. Int Orthop. 2017;41:611e618.
9. Philippot R, Adam P, Reckhaus M, et al. Prevention of dislocation in total hip revision surgery using a dual mobility design. Orthop Traumatol Surg Res. 2009;95:407e413.
10. Guyen O, Pibarot V, Vaz G, Chevillotte C, Bejui-Hugues J. Use of a dual mobility socket to manage total hip arthroplasty instability. Clin Orthop Relat Res. 2009;467:465e472.
11. Vasukutty NL, Middleton RG, Matthews EC, Young PS, Uzoigwe CE, Minhas TH. The double-mobility acetabular component in revision total hip replacement: the United Kingdom experience. J Bone Joint Surg Br. 2012;94:603e608.
12. Alberton GM, High WA, Morrey BF. Dislocation after revision total hip arthroplasty: an analysis of risk factors and treatment options. J Bone Joint Surg Am. 2002;84-A:1788e1792.
13. Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91: 128e133.
14. D'Antonio JA, Capello WN, Borden LS, et al. Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop Relat Res. 1989;243:126e137.
15. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20e32.
16. Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17e27.
17. Nilsdotter A, Bremander A. Measures of hip function and symptoms: Harris hip Score (HHS), hip disability and osteoarthritis outcome Score (HOOS), oxford hip Score (OHS), lequesne index of severity for osteoarthritis of the hip (LISOH), and American Academy of orthopedic surgeons (AAOS) hip and knee questionnaire. Arthritis Care Res. 2011;63:S200eS207.
18. Wegrzyn J, Tebaa E, Jacquel A, Carret JP, Bejui-Hugues J, Pibarot V. Can dual mobility cups prevent dislocation in all situations after revision total hip arthroplasty? J Arthroplasty. 2015;30:631e640.
19. Mohaddes M, Cnudde P, Rolfson O, Wall A, K€ arrholm J. Use of dual-mobility cup in revision hip arthroplasty reduces the risk for further dislocation: analysis of seven hundred and ninety one first-time revisions performed due to dislocation, reported to the Swedish Hip Arthroplasty Register. Int Orthop. 2017;41: 583e588.
20. Stucinskas J, Kalvaitis T, Smailys A, Robertsson O, Tarasevicius S. Comparison of dual mobility cup and other surgical constructs used for three hundred and sixty two first time hip revisions due to recurrent dislocations: five year results from Lithuanian arthroplasty register. Int Orthop. 2018;42:1015e1020.
21. Hartzler MA, Abdel MP, Sculco PK, Taunton MJ, Pagnano MW, Hanssen AD. Otto aufranc award: dual-mobility constructs in revision THA reduced dislocation, rerevision, and reoperation compared with large femoral heads. Clin Orthop Relat Res. 2018;476:293e301.
22. De Martino I, D'Apolito R, Soranoglou VG, Poultsides LA, Sculco PK, Sculco TP. Dislocation following total hip arthroplasty using dual mobility acetabular components: a systematic review. Bone Joint J. 2017;99-B:18e24.
23. Hamadouche M, Ropars M, Rodaix C, et al. Five to thirteen year results of a cemented dual mobility socket to treat recurrent dislocation. Int Orthop. 2017;41:513e519.
24. Wetters NG, Murray TG, Moric M, Sporer SM, Paprosky WG, Della Valle CJ. Risk factors for dislocation after revision total hip arthroplasty. Clin Orthop Relat Res. 2013;471:410e416.
25. Zywiel MG, Mustafa LH, Bonutti PM, Mont MA. Are abductor muscle quality and previous revision surgery predictors of constrained liner failure in hip arthroplasty? Int Orthop. 2011;35:797e802.
26. Ozden VE, Dikmen G, Beksac B, Tozun R. Dual-mobility bearings for patients with abductor-trochanteric complex insufficiency. Hip Int. 2018;28: 491e497.
27. Mertl P, Combes A, Leiber-Wackenheim F, Fessy MH, Girard J, Migaud H. Recurrence of dislocation following total hip arthroplasty revision using dual mobility cups was rare in 180 hips followed over 7 years. HSS J. 2012;8: 251e256.
28. Bauchu P, Bonnard O, Cypres A, Fiquet A, Girardin P, Noyer D. The dual- mobility POLARCUP: first results from a multicenter study. Orthopedics. 2008;31:1663e1672.
29. Darrith B, Courtney PM, Della Valle CJ. Outcomes of dual mobility components in total hip arthroplasty: a systematic review of the literature. Bone Joint J. 2018;100-B:11e19.