Measurement of the in-air output ratio for high-energy photon beams used in radiotherapy

To measure the in-air output ratio of a 15-MV photon beam using locally designed miniphantoms. Materials and methods: Columnar and brass miniphantoms were designed locally to accommodate 0.6 cc and 0.13 cc ionization chambers. The in-air output ratio (SC) was measured for square, rectangular, and wedged fields for 15 MV. The influences of the orientation of the miniphantom, miniphantom material, and chamber volume on SC and the collimator exchange effect were also studied. Results: The SC measurements ranged from 0.944 to 1.0321 for the studied field sizes. The orientation of the miniphantom had no influence on SC for the range of field sizes studied. The collimator exchange effect was found to be within 1.57%. The SC increased with wedged field sizes and wedge angles due to greater attenuation and production of low energy scatters. Conclusion: This study suggests the combination of a polymethylmethacrylate miniphantom for larger field sizes and a brass miniphantom for smaller field sizes when measuring SC in high-energy photon beams. The brass miniphantom with 0.6 cc and 0.13 cc ion chambers gives acceptable SC values for small field sizes. The chamber volume (0.6 cc or 0.13 cc) has no impact when used with a brass miniphantom.

Measurement of the in-air output ratio for high-energy photon beams used in radiotherapy

To measure the in-air output ratio of a 15-MV photon beam using locally designed miniphantoms. Materials and methods: Columnar and brass miniphantoms were designed locally to accommodate 0.6 cc and 0.13 cc ionization chambers. The in-air output ratio (SC) was measured for square, rectangular, and wedged fields for 15 MV. The influences of the orientation of the miniphantom, miniphantom material, and chamber volume on SC and the collimator exchange effect were also studied. Results: The SC measurements ranged from 0.944 to 1.0321 for the studied field sizes. The orientation of the miniphantom had no influence on SC for the range of field sizes studied. The collimator exchange effect was found to be within 1.57%. The SC increased with wedged field sizes and wedge angles due to greater attenuation and production of low energy scatters. Conclusion: This study suggests the combination of a polymethylmethacrylate miniphantom for larger field sizes and a brass miniphantom for smaller field sizes when measuring SC in high-energy photon beams. The brass miniphantom with 0.6 cc and 0.13 cc ion chambers gives acceptable SC values for small field sizes. The chamber volume (0.6 cc or 0.13 cc) has no impact when used with a brass miniphantom.

___

  • Frye DM, Paliwal BR, Thomadsen BR, Jursinic P.
  • Intercomparison of normalized head-scatter factor
  • measurement techniques. Med Phys 1995; 22: 249–53.
  • Zhu TC, Ahnesjö A, Lam KL, Li XA, Ma CM, Palta JR et al. Report of AAPM Therapy Physic Committee Task Group 74: In-air output ratio, SC, for megavoltage photon beams. Med Phys 2009; 36: 5261–91.
  • Jursinic PA. Measurement of head scatter factor of linear accelerators with columnar miniphantom. Med Phys 2006; 33: 1720–8.
  • Venselaar J, Heukelom S, Jager N, Mijnheer B J, van der Laarse, van Gasteren H et al. Effect of electron contamination on scatter correction factor for photon beam dosimetry. Med Phys 1999; 26: 2099–106.
  • Heukelom S, Lanson JH, Mijnheer BJ. Wedge factor constituents of higher energy photon beams: head and phantom scatter dose components. Radiother Oncol 1994; 32: 73–83.
  • Shih D, Li XA, Chu JC. Dynamic wedge versus physical wedge: a Monte Carlo study. Med Phys 2001; 28: 612–9.
  • Ding GX. An investigation of accelerator head scatter and output factor in air. Med Phys 2004; 31: 2527–33.
Turkish Journal of Medical Sciences-Cover
  • ISSN: 1300-0144
  • Yayın Aralığı: Yılda 6 Sayı
  • Yayıncı: TÜBİTAK
Sayıdaki Diğer Makaleler

The in vitro determination of genotoxicity in peripheral lymphocytes of welders exposed to fumes from metal arc welding

Ceylan ŞENER, Halil Erhan EROĞLU

No major impact of insertion/deletion polymorphism of the angiotensin-converting enzyme gene on thyroid-associated ophthalmopathy

Seydi OKUMUŞ, Şeniz DEMİRYÜREK, Mehtap ÖZKUR, Erol COŞKUN

The assessment of accessory mental foramina using cone-beam computed tomography

İbrahim Şevki BAYRAKDAR, Özkan MİLOĞLU, Hayati Murat AKGÜL, Mustafa GÖREGEN, İbrahim ERSOY

The correlation between metabolic syndrome and lower urinary tract symptoms in females

Cevdet Serkan GÖKKAYA, Cüneyt ÖZDEN, Binhan Kağan AKTAŞ, Hüseyin Cihan DEMİREL, Özer GÜZEL, Mehmet KARABAKAN, Akif Ersoy ERKMEN, Çetin Volkan ÖZTEKİN, Ali MEMİŞ

Do characteristics of seriously injured older adults differ from those of their younger counterparts in the emergency department?

Hüseyin ŞAHİN, Ayhan AKÖZ, Memet IŞIK, Mücahit EMET

Rapid identification of Aeromonas species in stool samples with chromogenic media and matrix-assisted laser desorption ionization-time of flight mass spectrometry: an institutional experience

Işın AKYAR, Simge CAN

Serum interleukin-8, CA-125 levels, neutrophil-to-lymphocyte ratios, and combined markers in the diagnosis of endometriosis

Cemile DAYANGAN SAYAN, Müzeyyen Gülnur ÖZAKŞİT, Esma SARIKAYA

JAK2 V617F mutation in Iranian patients with myeloproliferative neoplasms: clinical and laboratory findings

Behzad POOPAK, Majid FARSHDOUSTI HAGH, Najmaldin SAKI, Fazel ELAHI, Hamid REZVANI, Gelareh KHOSRAVIPOUR, Mohammad Ali JAHANGIRPOUR, Shirin BOLOURI, Tolou GOLKAR, Parviz FALAH

Is renal cell cancer stage migration valid throughout the world?

Emin Yener GÜLTEKİN, Esin YILDIZ, Semih AYAN, Esat KORĞALI, Hüseyin SAYGIN, Ziynet ÇINAR, Gökhan GÖKÇE

Effects of ethyl pyruvate administration on female rats' pyramidal cells of cornu ammonis after brain ischemia: a stereological and histopathological study

Orhan BAŞ, Ersan ODACI, Ayşe İKİNCİ