Effect of cycloplegia on refractive errors measured with three different refractometers in school-age children

Aim: To compare 3 different refractometers with and without cycloplegia and to examine whether the photorefractometer necessitates cycloplegia in the measurement of refractive errors. Materials and methods: Included in the study were 62 eyes of 31 pediatric patients. The refractive errors of all of the eyes were measured with and without cycloplegia using a table-top autorefractometer (Potec PRK-6000), hand-held autorefractometer (Nidek ARK-30), and photorefractometer (Plusoptix S08), respectively. The spherical power, cylindrical power, cylindrical axis, spherical equivalent, and interpupillary distance values obtained were statistically compared. Results: The average age of the patients was 10.03 ± 2.79 years. There were statistically significant differences between the cycloplegic and noncycloplegic spherical powers and the spherical equivalent values of each device. However, the response to cycloplegia was not significant for the cylindrical values. The Jackson cross-cylinder values at axis 0° and 45° (J0 and J45) of each device similarly was not significantly affected by the cycloplegia. Noncycloplegic spherical equivalent, cylindrical power, and J0 and J45 values measured with the Plusoptix S08 did not have a significant difference from the same values measured with the Potec PRK-6000 for cycloplegia. Conclusion: Accommodation has a prominent effect on the detection of refractive errors of school-age children. The photorefractometer method eliminates the need for cycloplegia in the detection of refractive errors of children from this age group. In the measurements performed with a hand-held autorefractometer, the tendency of myopia as a result of marked accommodation can be seen.

Effect of cycloplegia on refractive errors measured with three different refractometers in school-age children

Aim: To compare 3 different refractometers with and without cycloplegia and to examine whether the photorefractometer necessitates cycloplegia in the measurement of refractive errors. Materials and methods: Included in the study were 62 eyes of 31 pediatric patients. The refractive errors of all of the eyes were measured with and without cycloplegia using a table-top autorefractometer (Potec PRK-6000), hand-held autorefractometer (Nidek ARK-30), and photorefractometer (Plusoptix S08), respectively. The spherical power, cylindrical power, cylindrical axis, spherical equivalent, and interpupillary distance values obtained were statistically compared. Results: The average age of the patients was 10.03 ± 2.79 years. There were statistically significant differences between the cycloplegic and noncycloplegic spherical powers and the spherical equivalent values of each device. However, the response to cycloplegia was not significant for the cylindrical values. The Jackson cross-cylinder values at axis 0° and 45° (J0 and J45) of each device similarly was not significantly affected by the cycloplegia. Noncycloplegic spherical equivalent, cylindrical power, and J0 and J45 values measured with the Plusoptix S08 did not have a significant difference from the same values measured with the Potec PRK-6000 for cycloplegia. Conclusion: Accommodation has a prominent effect on the detection of refractive errors of school-age children. The photorefractometer method eliminates the need for cycloplegia in the detection of refractive errors of children from this age group. In the measurements performed with a hand-held autorefractometer, the tendency of myopia as a result of marked accommodation can be seen.

___

  • Schimitzek T, Haase W. Effi ciency of a video-autorefractometer used as a screening device for amblyogenic factors. Graefes Arch Clin Exp Ophthalmol 2002; 240: 710-6.
  • Mohindra I, Jacobson SG, Held R. Binocular visual form deprivation in human infants. Doc Ophthalmol 1983; 55: 237- 49.
  • Magramm I. Amblyopia: etiology, detection, and treatment. Pediatr Rev 1992; 13: 7-14.
  • Lithander J, Sjöstrand J. Anisometropic and strabismic amblyopia in the age group 2 years and above: a prospective study of the results of treatment. Br J Ophthalmol 1991; 75: 111-6.
  • Schimitzek T, Lagreze WA. Accuracy of a new photo- refractometer in young and adult patients. Graefes Arch Clin Exp Ophthalmol 2005; 243: 637-45.
  • Abrahamsson M, Ohlsson J, Björndahl M, Abrahamsson H. Clinical evaluation of an eccentric infrared photorefractor: the PowerRefractor. Acta Ophthalmol Scand 2003; 81: 605-10.
  • Egashira SM, Kish LL, Twelker JD, Mutti DO, Zadnik K, Adams AJ. Comparison of cyclopentolate versus tropicamide cycloplegia in children. Optom Vis Sci 1993; 70: 1019-26.
  • Fotedar R, Rochtchina E, Morgan I, Wang JJ, Mitchell P, Rose KA. Necessity of cycloplegia for assessing refractive error in 12-year-old children: a population-based study. Am J Ophthalmol 2007; 144: 307-9.
  • Holland BJ, Siderov J. Repeatability of measurements of interpupillary distance. Ophthalmic Physiol Opt 1999; 19: 74- 8.
  • Hunt OA, Wolff sohn JS, Gilmartin B. Evaluation of the measurement of refractive error by the PowerRefractor: a remote, continuous and binocular measurement system of oculomotor function. Br J Ophthalmol 2003; 87: 1504-8.
  • Cordonnier M, Dramaix M. Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatism. Br J Ophthalmol 1999; 83: 157-61.
  • Paff T, Oudesluys-Murphy AM, Wolterbeek R, Swart-van den Berg M, de Nie JM, Tijssen E et al. Screening for refractive errors in children: the plusoptiX S08 and the Retinomax K-plus2 performed by a lay screener compared to cycloplegic retinoscopy. J AAPOS 2010; 14: 478-83.
  • Th ibos LN, Wheeler W, Horner D. Power vectors: an application of Fourier analysis to the description and statistical analysis of refractive error. Optom Vis Sci 1997; 74: 367-75.
  • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-10.
  • Wesson MD, Mann KR, Bray NW. A comparison of cycloplegic refraction to the near retinoscopy technique for refractive error determination. J Am Optom Assoc 1990; 61: 680-4.
  • Saunders KJ, Westall CA. Comparison between near retinoscopy and cycloplegic retinoscopy in the refraction of infants and children. Optom Vis Sci 1992; 69: 615-22.
  • Borghi RA, Rouse MW. Comparison of refraction obtained by “near retinoscopy” and retinoscopy under cycloplegia. Am J Optom Physiol Opt 1985; 62: 169-72.
  • Gettes BC, Belmont O. Tropicamide: comparative cycloplegic eff ects. Arch Ophthalmol 1961; 66: 336-40.
  • Stolzar IH. A new group of cycloplegic drugs. Am J Ophthalmol 1953; 36: 110-2.
  • Mutti DO, Zadnik K, Egashira S, Kish L, Twelker JD, Adams AJ. Th e eff ect of cycloplegia on measurement of the ocular components. Invest Ophthalmol Vis Sci 1994; 35: 515-27.
  • Twelker JD, Mutti DO. Retinoscopy in infants using a near noncycloplegic technique, cycloplegia with tropicamide 1%, and cycloplegia with cyclopentolate 1%. Optom Vis Sci 2001; 78: 215-22.
  • Th ompson AM, Li T, Peck LB, Howland HC, Counts R, Bobier WR. Accuracy and precision of the Tomey ViVA infrared photorefractor. Optom Vis Sci 1996; 73: 644-652.
  • Wesemann W, Norcia AM, Allen D. Th eory of eccentric photorefraction (photoretinoscopy): astigmatic eyes. J Opt Soc Am A 1991; 8: 2038-47.
  • Choi M, Weiss S, Schaeff el F, Seidemann A, Howland HC, Wilhelm B et al. Laboratory, clinical, and kindergarten test of a new eccentric infrared photorefractor (PowerRefractor). Optom Vis Sci 2000; 77: 537-48.
  • Allen PM, Radhakrishnan H, O’Leary DJ. Repeatability and validity of the PowerRefractor and the Nidek AR600-A in an adult population with healthy eyes. Optom Vis Sci 2003; 80: 245-51.
  • Gekeler F, Schaeff el F, Howland HC, Wattam-Bell J. Measurement of astigmatism by automated infrared photoretinoscopy. Optom Vis Sci 1997; 74: 472-82.
  • Harvey EM, Miller JM, Wagner LK, Dobson V. Reproducibility and accuracy of measurements with a hand held autorefractor in children. Br J Ophthalmol 1997; 81: 941-8.
  • Filipović T. Changes in the interpupillary distance (IPD) with ages and its eff ect on the near convergence/distance (NC/D) ratio. Coll Antropol 2003; 27: 723-7.
  • Alp MN, Atilla H, Erkam N. A comparative study of refraction in emmetropic eyes with automated refractometer and retinoscopy. T Klin J Ophthalmol 1998; 7: 178-83.
Turkish Journal of Medical Sciences-Cover
  • ISSN: 1300-0144
  • Yayın Aralığı: Yılda 6 Sayı
  • Yayıncı: TÜBİTAK
Sayıdaki Diğer Makaleler

The effect of N-acetylcysteine on growth and biofi lm formation in Staphylococcus epidermidis strains

Sahra KIRMUSAOĞLU, Seyhun YURDUGÜL, M. Esra KOÇOĞLU

The impact of nephrometry score on partial nephrectomy rates and survival

Ali Barbaros BAŞESKİOĞLU, Yunus Saim Cavit CAN, Aydın YENİLMEZ, Coşkun KAYA

Ileosigmoid knotting in pregnancy

Sabri Selçuk ATAMANALP

Isı Şok Protein İnhibitörü KNK437’nin Endotel Hücrelerinde Anjiyogenez Üzerine Etkisi

Mehmet ŞAHİN, Emel ŞAHİN, Saadet GÜMÜŞLÜ

Does the treatment of obstructive sleep apnea syndrome improve lower urinary tract symptoms?

Öner ODABAŞ, Metin YIĞMAN, Cavit CEYLAN

The role of quantitative D-dimer levels in the follow-up and differential diagnosis of pulmonary thromboembolism and community-acquired pneumonia

Müge AYDOĞDU, Numan Nadir EKİM, Ayhan VAROL, Nurdan KÖKTÜRK, Hatice KILIÇ

Comparison between intravenous and thoracic epidural analgesia in single coronary artery bypass graft surgery

Ali Cem YORGANCIOĞLU, Serdar GÜNAYDIN, Nevriye SALMAN, Halil İbrahim UÇAR, Fatih Tanzer SERTER

Iodine deficiency in pregnancy and in women of reproductive age in Erzurum, Turkey

Metin İNGEÇ, Semra ÇETİNKAYA, Kadir ÇETİNKAYA, İbrahim KAPLAN

Comparison of anesthesia with sevofl urane-N2O and midazolam-remifentanil in low-birth-weight premature infants undergoing diode laser photocoagulation

Levent ÖZTÜRK, Eyüp HORASANLI, Abdulkadir BUT, Müge ARIKAN, Meltem TABUK, Bilge ASLAN

Cytotoxic effect of cigarette smoke condensate on mice and rat mesenchymal stem cells and HeLa cells

Aftab AHMAD, Abdul Rauf SHAKOORI