Characterization of Klebsiella isolates by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) and determination of antimicrobial resistance with VITEK 2 advanced expert system (AES)

Background/aim: The purpose of the study was to evaluate the performance of the VITEK mass spectrometry (MS) (bioMérieux, France) system for the identification of Klebsiella spp. isolated from different sources. Moreover, while assessing the ability of the VITEK 2 automated expert system (AES) to recognize antimicrobial resistance patterns, the researchers have extended the study to compare VITEK 2 with the routine antimicrobial susceptibility testing method. Materials and methods: This study tested 51 Klebsiella spp. isolates that were isolated from environmental examples and clinical examples. Results of conventional methods and the matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) MS were compared. Then, any differing results were compared against a reference 16S rRNA gene sequence, and when indicated, a recA sequencing analysis was done. Results: VITEK MS correctly identified 100% of the Klebsiella spp. isolates. There were two K. oxytoca isolates incorrectly identified to the species level with conventional methods according to the 16S rRNA gene sequencing analysis. In addition, a VITEK 2 AST-N261 card was used for the detection of extended spectrum beta-lactamases (ESBL). Using the VITEK 2 AES, ESBL positivity was found at the rate of 16.3% whereas this rate was 4.08% using the disk diffusion method. Conclusion: MALDI-TOF MS is a rapid and accurate method for the identification of Klebsiella spp. Moreover, the bioMérieux AES provides a useful laboratory tool for the interpretation of susceptibility results.

Characterization of Klebsiella isolates by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) and determination of antimicrobial resistance with VITEK 2 advanced expert system (AES)

Background/aim: The purpose of the study was to evaluate the performance of the VITEK mass spectrometry (MS) (bioMérieux, France) system for the identification of Klebsiella spp. isolated from different sources. Moreover, while assessing the ability of the VITEK 2 automated expert system (AES) to recognize antimicrobial resistance patterns, the researchers have extended the study to compare VITEK 2 with the routine antimicrobial susceptibility testing method. Materials and methods: This study tested 51 Klebsiella spp. isolates that were isolated from environmental examples and clinical examples. Results of conventional methods and the matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) MS were compared. Then, any differing results were compared against a reference 16S rRNA gene sequence, and when indicated, a recA sequencing analysis was done. Results: VITEK MS correctly identified 100% of the Klebsiella spp. isolates. There were two K. oxytoca isolates incorrectly identified to the species level with conventional methods according to the 16S rRNA gene sequencing analysis. In addition, a VITEK 2 AST-N261 card was used for the detection of extended spectrum beta-lactamases (ESBL). Using the VITEK 2 AES, ESBL positivity was found at the rate of 16.3% whereas this rate was 4.08% using the disk diffusion method. Conclusion: MALDI-TOF MS is a rapid and accurate method for the identification of Klebsiella spp. Moreover, the bioMérieux AES provides a useful laboratory tool for the interpretation of susceptibility results.

___

  • Fabri G, Panico M, Dallolio L, Suzzi R, Ciccia M, Sandri F, Farruggia P. Outbreak of ampicillin/piperacillin-resistant Klebsiella pneumoniae in a neonatal intensive care unit (NICU): investigation and control measures. Int J Environ Res Public Health 2013; 10: 808–815.
  • Gastmeier P, Loui A, Stamm-Balderjahn S, Hansen S, Zuschneid I, Sohr D, Behnke M, Obladen M, Vonberg RP, Ruden H. Outbreaks in neonatal intensive care units – they are not like others. Am J Infect Control 2007; 35: 172–176.
  • Casolari C, Pecorari M, Fabio G, Cattani S, Venturelli C, Piccinini L, Tamassia MG, Gennari W, Sabbatini AM, Leporati G et al. A simultaneous outbreak of Serratia marcescens and Klebsiella pneumoniae in a neonatal intensive care unit. J Hosp Infect 2005; 61: 312–320.
  • Berrazeg M, Diene SM, Drissi M, Kempf M, Richet H, Landraud L, Rolain J-M. Biotyping of multidrug-resistant Klebsiella pneumoniae clinical isolates from France and Algeria using MALDI-TOF MS. Plos One 2013; 8: e61428.
  • Kumar A, Chakraborti S, Joshi P, Chakrabarti P, Chakraborty R. A multiple antibiotic and serum resistant oligotrophic strain, Klebsiella pneumoniae MB45 having novel dfrA30, is sensitive to ZnO QDs. Ann Clin Microbiol Antimicrob 2011; 10: 19.
  • Brisse S, Duijkeren E. Identification and antimicrobial susceptibility of 100 Klebsiella animal clinical isolates. Vet Microbiol 2005; 105: 307–312.
  • Struve C, Krogfelt KA. Pathogenic potential of environmental Klebsiella pneumoniae isolates. Environ Microbiol 2004; 6: 584–590.
  • Gupta A, Della-Latta P, Todd B, San Gabriel P, Haas J, Wu F, Rubenstein D, Saiman L. Outbreak of extended-spectrum beta-lactamase-producing Klebsiella pneumoniae in a neonatal intensive care unit linked to artificial nails. Infect Control Hosp Epidemiol 2004; 25: 210–215.
  • Pessoa-Silva CL, Meurer Moreira B, Câmara Almeida V, Flannery B, Almeida Lins MC, Mello Sampaio, JL, Martins Teixeira L, Vaz Miranda LE, Riley LW, Gerberding JL. Extended-spectrum beta-lactamase-producing Klebsiella pneumoniae in a neonatal intensive care unit: risk factors for infection and colonization. J Hosp Infect 2003; 53: 198–206.
  • Wang YR, Qian C, Hui CS, Qin LF. Characterization of Staphylococcus aureus isolated from clinical specimens by matrix assisted laser desorption/ionization time-of-flight mass spectrometry. Biomed Environ Sci 2013; 26: 430–436.
  • Turner PJ. Extended-spectrum beta-lactamases, Clin Infect Dis 2005; 41: S273–S275.
  • Çetinkaya E, Ayhan K. Molecular technics used in microbiology. Karaelmas Sci Eng J 2012; 2: 53–62 (article in Turkish).
  • Yılmaz E, Uraz G. Detection of extended spectrum beta lactamase in Klebsiella pneumoniae and Klebsiella oxytoca bacteria with combined disk method. Sakarya University Sci J 2011; 15:1 41–45 (article in Turkish).
  • Akçam FZ, Gönen I, Kaya O, Yaylı G. Investigating sensitivity to beta-lactam antibiotics and ESBL frequency in hospital infection factor enterobacteria. Klimik J 2004; 17: 47–49.
  • Sanders CC, Peyret M, Moland ES, Cavalieri SJ, Shubert C, Thomson KS, Boeufgras JM, Sanders WE Jr. Potential impact of the VITEK 2 system and the advanced expert system on the clinical laboratory of a university-based hospital. J Clin Microbiol 2001; 39: 2379–2385.
  • Valenza G, Müller S, Schmitt C, Turnwald D, Lam TT, Frosch M, Abele-Horn M, Pfeifer Y. Evaluation of the VITEK 2 AST-N111 card for detection of extended-spectrum beta- lactamases (ESBLs) in Escherichia coli, Klebsiella pneumoniae, and Klebsiella oxytoca compared to ESBL Etest and combination disc methods. Eur J Clin Microbiol Infect Dis 2011; 30: 869– 872.
  • VITEK MS User Manual. France: bioMérieux; 2011.
  • Cesarini S, Bevivino A, Tabacchioni S, Chiarini L, Dalmastri C. RecA gene sequence and multilocus sequence typing for species-level resolution of Burkholderia cepecia complex isolates. Lett Appl Microbiol 2009; 49: 580–588.
  • Clinical and Laboratory Standards Institute (CLSI). Performance Standards for Antimicrobial Susceptibility Testing, 21st Informational Supplement M100-S20. Wayne, PA, USA: CLSI; 2010.
  • Yssouf A, Socolovschi C, Flaudrops C, Ndiath MO, Sogoufara S, Dehecq J-S, Lacour G, Berenger J-M, Sokhna S, Raoult D et al. Matrix-assisted laser desorption ionization – time of flight mass spectrometry: an emerging tool for the rapid identification of mosquito vectors. Plos One 2013, 8: e72380.
  • Manji R, Bythrow M, Branda JA, Burnham C-AD, Ferraro MJ, Garner OB, Jenneman R, Lewinski MA, Mochon AB, Procop GW et al. Multi-center evaluation of the VITEK MS system for mass spectrometric identification of non-Enterobacteriaceae gram-negative bacilli. Eur J Clin Microbiol Infect Dis 2014; 33: 337–346.
  • Richter SS, Sercia L, Branda JA, Burnham CAD, Bythrow M, Ferrano MJ, Garner OB, Ginocchio CC, Jenneman R, Lewinski MA et al. Identification of Enterobacteriaceae by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry using the VITEK MS system. Eur J Clin Microbiol Infect Dis 2013; 32: 1571–1578.
  • Saffert RT, Cunningham SA, Ihde SM, Jobe KE, Mandrekar J, Patel R. Comparison of Bruker Biotyper matrix-assisted laser desorption ionization time of flight mass spectrometer to BD Phoenix automated microbiology system for identification of gram-negative bacilli. J Clin Microbiol 2011; 49: 887–892.
  • Barry J, Brown A, Ensor V, Lakhani U, Petts D, Warren C, Winstanley T. Comparative evaluation of the VITEK 2 advanced expert system (AES) in five UK hospitals. J Antimicrob Chemother 2003; 51: 1191–1202.
  • Cantón R, Pérez-Vázquez M, Oliver A, Coque TM, Loza E, Ponz F, Baquero F. Validation of the VITEK2 and the advanced expert system with a collection of Enterobacteriaceae harbouring extended spectrum or inhibitor resistant beta- lactamases. Diagn Microbiol Infect Dis 2001; 41: 65–70.
  • Leclercq R, Nicolas-Chanoine MH, Nordmann P, Philippon A, Marchais P, Buu-Ho A, Chardon H, Dabernat H, Doucet- Populaine F, Grasmick C et al. Multicentre evaluation of an automated system using selected bacteria that harbour challenging and clinically relevant mechanisms of resistance to antibiotics. Eur J Clin Microbiol Infect Dis 2001; 20: 626–635.
  • Nyberg SD, Muerman O, Jalava J, Rantakoko-Jalava K. Evaluation of detection of extended-spectrum beta lactamases among Escherichia coli and Klebsiella spp. isolates by VITEK 2 AST-N029 compared to the agar dilution and disk diffusion methods. Scand J Infect Dis 2008; 40: 355–362.
  • Sorlozano A, Gutierrez J, Piedrola G, Soto MJ. Acceptable performance of VITEK 2 system to detect extended spectrum beta-lactamases in clinical isolates of Escherichia coli: a comparative study of phenotypic commercial methods and NCCLS guidelines. Diagn Microbiol Infect Dis 2005; 51:191– 193.
  • Livermore DM, Struelens M, Amorim J, Baquero F, Bille J, Canton R, Henning S, Gatermann S, Marchese A, Mittermayer H et al. Multicentre evaluation of the VITEK 2 advanced expert system for interpretive reading of antimicrobial resistance tests. J Antimicrob Chemother 2002; 49: 289–300.
  • Blondel-Hill E, Hetchler C, Andrews D, Lapointe L. Evaluation of VITEK 2 for analysis of Enterobacteriaceae using the advanced expert system (AES) versus interpretive susceptibility guidelines used at Dynacare Kapser Medical Laboratories, Edmonton, Alberta. Clin Microbiol Infect 2003; 9: 1091–1103.
  • Numanovic F, Hukic M, Delibegovic Z, Tihic N, Pasic S, Gegic M. Comparison of double disk synergy test, VITEK 2 and Check-MDR CT 102 for detection of ESBL producing isolates. Acta Media Academica 2013; 42: 15–24.
Turkish Journal of Medical Sciences-Cover
  • ISSN: 1300-0144
  • Yayın Aralığı: Yılda 6 Sayı
  • Yayıncı: TÜBİTAK
Sayıdaki Diğer Makaleler

The effectiveness of ultrasound treatment for the management of kneeosteoarthritis: a randomized, placebo-controlled, double-blind study

Serap Kapci YILDIZ, Feyza Ünlü ÖZKAN, İlknur AKTAŞ, Ayşe Duygu ŞİLTE, Meryem Yilmaz KAYSIN, Naciye Bilgin BADUR

Molecular characteristics of extended spectrum β-lactamase and carbapenemase genes carried by carbapenem-resistant Enterobacter cloacae in a Chinese university hospital

Lanying CUI, Jinying ZHAO, Juan LU

Restless legs syndrome in multiple sclerosis

Mine Hayriye SORGUN, Zerin AKSUN, Yahya Burak ATALAY, CANAN YÜCESAN

Factors affecting daily instrumental activities of the elderly

Ayfer ŞAHİN, Oğuz TEKİN, Sevsen CEBECİ, Bünyamin IŞIK, Adem ÖZKARA, Rabia KAHVECİ, Metin CANBAL, Şeyda KUNT, İrfan ŞENCAN

Molecular characteristics of extended spectrum ß-lactamase and carbapenemase genes carried by carbapenem-resistant Enterobacter cloacae in a Chinese university hospital

Lanying CUI, Jingying ZHAO, Juan LU

The evaluation of hypermetabolic mediastinal hilar lymph nodes determined by PET/CT in pulmonary and extrapulmonary malignancies: correlation with EBUS-TBNA

Hatice Canan HASANOĞLU, Elif ÖZDEMİR, Şeyda TÜRKÖLMEZ, Nilüfer YILDIRIM, Aysegül ŞENTÜRK, Zuhal KANDEMİR, Mutlay KESKİN

PCR investigation of Panton-Valentine leukocidin, enterotoxin, exfoliative toxin, and agr genes in Staphylococcus aureus strains isolated from psoriasis patients

Jülide Sedef GÖÇMEN, Mukadder KOÇAK, Neriman ŞAHİNER, Zeynep Ceren KARAHAN

Decreasing the need for mechanical ventilation after surgery forretinopathy of prematurity: sedoanalgesia vs. general anesthesia

Ayşe ÜLGEY, İşin GÜNEŞ, Adnan BAYRAM, Recep AKSU, Cihangir BİÇER, Fatih UĞUR, Çağatay KARACA, Adem BOYACI

The evaluation of hypermetabolic mediastinal?hilar lymph nodes determined byPET/CT in pulmonary and extrapulmonary malignancies: correlation with EBUS-TBNA

ZUHAL KANDEMİR, AYŞEGÜL ŞENTÜRK, ELİF ÖZDEMİR, NİLÜFER YILDIRIM, HATİCE CANAN HASANOĞLU, MUTLAY KESKİN, ŞEYDA TÜRKÖLMEZ

An anatomical study of the supratrochlear foramen of the Jining population

JING LI, QINGHUA MAO, WENYUAN LI, XIAOYAN LI