Dosimetric Comparison of Anisotropic Analytical Algorithm and Acuros XB in Stereotactic Body Radiotherapy and Effect of Calculation Grid Size

Dosimetric Comparison of Anisotropic Analytical Algorithm and Acuros XB in Stereotactic Body Radiotherapy and Effect of Calculation Grid Size

This study aimed to determine the dose differences between Acuros XB (AXB) and anisotropic analytical algorithm (AAA) for patients with liver cancer who underwent stereotactic body radiotherapy (SBRT) and to investigate the dose-related effect of dose calculation grid size (CGS).METHODSSBRT treatment was planned for 10 patients with liver cancer using 1-mm and 2.5-mm CGS with AAA and AXB algorithms with 10 MV flattening filter-free (FFF) rays in Varian Trubeam STx.RESULTSWhen AAA 1-mm CGS plans and AAA 2.5-mm CGS plans were compared, 4% difference was observed; when AXB 1-mm CGS plans and AXB 2.5-mm CGS plans were compared, 1% difference was found. No significant difference was found between plans with AAA 1-mm CGS and plans with AXB 1-mm CGS (p>0.05). On the other hand, there was a significant difference between plans with AAA 2.5mm CGS and plans with AXB 2.5-mm CGS (p

___

  • 1. Fraass BA, Smathers J, Deye J. Summary and recommendations of a National Cancer Institute workshop on issues limiting the clinical use of Monte Carlo dose calculation algorithms for megavoltage external beam radiation therapy. Med Phys 2003;30(12):3206-16.
  • 2. Vassiliev ON, Wareing TA, McGhee J, Failla G, Salehpour MR, Mourtada F. Validation of a new gridbased Boltzmann equation solver for dose calculation in radiotherapy with photon beams. Phys Med Biol 2010;55(3):581-98.
  • 3. Bush K, Gagne IM, Zavgorodni S, Ansbacher W, Beckham W. Dosimetric validation of Acuros XB with Monte Carlo methods for photon dose calculations. Med Phys 2011;38(4):2208-21.
  • 4. Hoffmann L, Jørgensen MB, Muren LP, Petersen JB. Clinical validation of the Acuros XB photon dose calculation algorithm, a grid-based Boltzmann equation solver. Acta Oncol 2012;51(3):376-85.
  • 5. Tillikainen L, Helminen H, Torsti T, Siljamäki S, Alakuijala J, Pyyry J, et al. A 3D pencil-beam-based superposition algorithm for photon dose calculation in heterogeneous media. Phys Med Biol 2008;53(14):3821-39.
  • 6. Tillikainen L, Siljamäki S, Helminen H, Alakuijala J, Pyyry J. Determination of parameters for a multiplesource model of megavoltage photon beams using optimization methods. Phys Med Biol 2007;52(5):1441- 67.
  • 7. Han T, Mikell JK, Salehpour M, Mourtada F. Dosimetric comparison of Acuros XB deterministic radiation transport method with Monte Carlo and model-based convolution methods in heterogeneous media. Med Phys 2011;38(5):2651-64.
  • 8. Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media. Radiat Oncol 2011;6(1):82.
  • 9. Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. On the dosimetric impact of inhomogeneity management in the Acuros XB algorithm for breast treatment. Radiat Oncol 2011;6:103.
  • 10.Robinson D. Inhomogeneity correction and the analytic anisotropic algorithm. J Appl Clin Med Phys 2008;9(2):2786.
  • 11.Breitman K, Rathee S, Newcomb C, Murray B, Robinson D, Field C, et al. Experimental validation of the Eclipse AAA algorithm. J Appl Clin Med Phys 2007;8(2):76-92.
  • 12.Van Esch A, Tillikainen L, Pyykkonen J, Tenhunen M,Helminen H, Siljamäki S, et al. Testing of the analytical anisotropic algorithm for photon dose calculation. Med Phys 2006;33(11):4130-48.
  • 13.Fogliata A, Nicolini G, Clivio A, Vanetti E, Cozzi L. Critical appraisal of Acuros XB and Anisotropic Analytic Algorithm dose calculation in advanced nonsmall-cell lung cancer treatments. Int J Radiat Oncol Biol Phys 2012;83(5):1587-95.
  • 14.Kan MW, Leung LH, Yu PK. Verification and dosimetric impact of Acuros XB algorithm on intensity modulated stereotactic radiotherapy for locally persistent nasopharyngeal carcinoma. Med Phys 2012;39(8):4705-14.
  • 15.Chung H, Jin H, Palta J, Suh TS, Kim S. Dose variations with varying calculation grid size in head and neck IMRT. Phys Med Biol 2006;51(19):4841-56.
  • 16.Mittauer K, Lu B, Yan G, Kahler D, Gopal A, Amdur R, et al. A study of IMRT planning parameters on planning efficiency, delivery efficiency, and plan quality. Med Phys 2013;40(6):061704.
  • 17.Ong CL, Cuijpers JP, Senan S, Slotman BJ, Verbakel WF. Impact of the calculation resolution of AAA for small fields and RapidArc treatment plans. Med Phys 2011;38(8):4471-9.
  • 18.Huang B, Wu L, Lin P, Chen C. Dose calculation ofAcuros XB and Anisotropic Analytical Algorithm in lung stereotactic body radiotherapy treatment with flattening filter free beams and the potential role of calculation grid size. Radiat Oncol 2015;10:53.