A comparative study between capillary and venous blood glucose levels of type 2 diabetes mellitus patients in intensive care units

Blood glucose estimation is required to be done rapidly or even on hourly basis for effective dosing of insulin in diabetic patients admitted in intensive care units (ICUs). Glucose test strip precision is usually considered clinically acceptable if the difference from the reference value (lab method) does not exceed 15%. With an aim to determine the correlation of capillary blood glucose versus venous blood glucose in hyperglycemic type 2 diabetes mellitus patients admitted in ICU, a study was conducted in a large tertiary care hospital. This study was a cross sectional, descriptive study which was carried out in a large tertiary care teaching hospital in Pune, from 01 Nov 2015 to 31 Oct 2016. After excluding all the confounding factors, 65 patients were included in this study. The laboratory results and glucose strip results were tabulated in a master chart as sample 1(capillary/venous- time of admission), sample 2(capillary /venous – 24 hrs), sample 3 (capillary /venous – 48 hrs) and also segregated into less than 250 mg/dl and more than 250 mg/dl for determining the correlation and agreement between the capillary and venous blood glucose levels. A good correlation existed between the CBG and VBG of sample 1 (R2 =0.995, p

___

Yaraghi A, Mood NE, Dolatabadi LK. Comparison of capillary and venous blood glucose levels using glucometer and laboratory blood glucose level in poisoned patients being in coma. Adv Biomed Res. 2015;23(4):247.

Juneja D, Pandey R, Singh O. Comparison between arterial and capillary blood glucose monitoring in patients with shock. Eur J Intern Med. 2011;22(3):241-4.

Friedri JO, Chant C, Adhikari NK. Does intensive insulin therapy reduce mortality in critically ill surgical patients? A reanalysis of meta analytic data. Crit Cacre. 2010;14(5)324.

Adnana M, Imamb F, Shabbira I, et al. Correlation between capillary and venous blood glucose levels in diabetic patients. Asian Biomedicine.2015;55-9.

Patel N, Patel K. A comparative study of venous and capillary blood glucose levels by different method GCSMC J Med Sci Vol (IV) No (I) January-June 2015.

Freckmann G, Schmid C, Baumstark A, et al. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012;6(5):1060-75.

Aghakachoei S, Zargarzadeh AH, Amini M. Comparison of blood glucose values using two glucose meters and standard laboratory method in hospitalized patients in a teaching hospital. J Pharm Care. 2014;2:15-21.

Kotwal N, Pandit A. Variability of capillary blood glucose monitoring measured on home glucose monitoring devices. Indian J Endocrinol Metab. 2012;16(Suppl 2)S248-51.

Kermani SK, Khatony A, Jalali R, et al. Accuracy and Precision of Measured Blood Sugar Values by three Glucometers Compared to the Standard Technique. J Clin Disqn Res.2017,11(4);OC05-OC08.

Van den Berghe G, Schetz M, Vlasselaers D, et al. Clinical review: Intensive Insulin therapy in critically Ill patients: NICE-SUGAR or Leuven blood glucose target? J. Clin. Endocrinol. Metab. 2009;94(9):3163-70.

Yang C, Chang C, Lin J. Department of food science national pingtung university of science and technology pingtung, taiwan, A comparison between venous and finger-prick blood sampling on values of blood glucose. IPCBEE vol.39 (2012).

International Organization for Standardization: In vitro diagnostic test systems–requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO report 15197. Geneva, Switzerland, International Organization for Standardization. 2003.

American Diabetes Association. Standards of medical care in diabetes-2012. Diabetes Care. 2012;35(Suppl 1):11-63.
Medicine Science-Cover
  • ISSN: 2147-0634
  • Yayın Aralığı: Yılda 4 Sayı
  • Başlangıç: 2012
  • Yayıncı: Effect Publishing Agency ( EPA )