Analyses of Surface Dose from High Energy Photon Beams for Different Clinical Setup Parameters

Surface and build-up region doses of 6, 15 and 25 MV photon beams were investigated for several clinical setup parameters. Evaluated setups included open fields, physical wedge (PW) fields, virtual wedge (VW) fields and blocked fields. The effects of field size, acrylic block tray, source-to-surface distance (SSD) and off-axis distance on surface dose were determined for each of these setups. Siemens Mevatron MD2 (6-15 MV) and Mevatron-KD2 (6-25 MV) linear accelerators were used to measure both the surface and build-up region doses. The surface dose increased as field size increased in open fields. There was no significant difference between 6 MV photon beam surface dose values for both machines. The surface dose for PW fields was lower than the dose for an open field, but higher in the case of large fields and higher degree wedges. The VW field surface doses were higher than the dose for PW and quite similar to the open fields. With the use of an acrylic block tray, the surface dose increased for all field sizes, but the increase was dominant for large fields. The surface dose for blocked fields was lower than the dose for open fields. As SSD decreased, the surface dose increased, and this effect was dominant, especially in larger field sizes with an acrylic block tray.
Anahtar Kelimeler:

skin dose, wedge, blocked field

Analyses of Surface Dose from High Energy Photon Beams for Different Clinical Setup Parameters

Surface and build-up region doses of 6, 15 and 25 MV photon beams were investigated for several clinical setup parameters. Evaluated setups included open fields, physical wedge (PW) fields, virtual wedge (VW) fields and blocked fields. The effects of field size, acrylic block tray, source-to-surface distance (SSD) and off-axis distance on surface dose were determined for each of these setups. Siemens Mevatron MD2 (6-15 MV) and Mevatron-KD2 (6-25 MV) linear accelerators were used to measure both the surface and build-up region doses. The surface dose increased as field size increased in open fields. There was no significant difference between 6 MV photon beam surface dose values for both machines. The surface dose for PW fields was lower than the dose for an open field, but higher in the case of large fields and higher degree wedges. The VW field surface doses were higher than the dose for PW and quite similar to the open fields. With the use of an acrylic block tray, the surface dose increased for all field sizes, but the increase was dominant for large fields. The surface dose for blocked fields was lower than the dose for open fields. As SSD decreased, the surface dose increased, and this effect was dominant, especially in larger field sizes with an acrylic block tray.
Turkish Journal of Medical Sciences-Cover
  • ISSN: 1300-0144
  • Yayın Aralığı: Yılda 6 Sayı
  • Yayıncı: TÜBİTAK
Sayıdaki Diğer Makaleler

Radioimmune Imaging of Diabetic Foot Infection -Tc-99m-Labelled Antigranulocyte Antibody in Combination with Tc-99m-Methylene Diphosphonate Bone Scintigraphy

Hripsime Artjun BOHCHELIAN

The Effects of Pentoxifylline in Experimental Nerve Injury

Süleyman BAYKAL

Prevalence of Asthma and Asthma-Like and Allergic Symptoms in the Urban Adult Population of Elazığ

Tuncer TUĞ, Esra TUĞ, Yasemin AÇIK

IFN-gamma-mediated transcriptional induction of the IDO (Indolamine 2,3-dioxygenase) gene requires activity of p68/PKR protein kinase

Nilüfer Gülmen İMİR, Osman Nidai ÖZEŞ, Tekinalp GELEN, Mehmet ÖĞÜŞ, Seyda KARAVELİ, Ayşe Gül NASIRCILAR, Hakan AKÇA

Genetic Polymorphisms in Steroid Hormone Metabolizing Enzymes in Human Breast Cancer

Neslihan AYGÜN KOCABAŞ

A Simple Comparison Between Simultaneous Kidney-Pancreas (SKPtx) and Kidney-Alone (Ktx) Transplants in Type I Diabetic Nephropaty

Tanju ACAR, Arshad MAHMOOD

The Rupture of a Large Isolated Iliac Artery Aneurysm

Sami KARAPOLAT, Yahya ÜNLÜ, Ünsal VURAL, Münacettin CEVİZ, Hikmet KOÇAK

Detection of Human Papillomavirus Infection in Esophageal Carcinomas by the Histopathological Method and Polymerase Chain Reaction Technique

İlhami KİKİ, Mehmet GÜNDOĞDU, Cemal GÜNDOĞDU, Fevzi POLAT

Radioimmune imaging of diabetic foot infection Tc-99m-labelled antigranulocyte antibody in combination with Tc-99m-methylene diphosphonate bone scintigraphy

Hripsime Artjun BOHCHELIAN, Anelia Dimitrova KLISAROVA, Lidia Asenova KOEVA

Pedodontic Complete Dentures

Mehmet DALKIZ, Bedri BEYDEMİR