Interobserver Reproducibility of Intracranial Anatomy Assessment During Second Trimester Sonographic Scan

Interobserver Reproducibility of Intracranial Anatomy Assessment During Second Trimester Sonographic Scan

To evaluate the reproducibility of mid-trimester intracranial anatomy assessment. Women undergoing mid-trimester scan (between 18th to 25th gestational weeks) for fetal anatomy assessment were included in the study. Measurements of lateral ventricle (LV), cisterna magna (CM) and transverse cerebellar diameter (TCD) were obtained for reproducibility analyses. Inter-observer reproducibility analysis was made with correlation coefficients. A total of 162 women were included in the analysis while one woman was excluded from the study due to diagnosis of vermian agenesis. Correlation coefficient (CC) of cisterna magna measurements have shown weak to moderate interclass correlation (r=0.28, P=0.001). Measurements of lateral ventricle have shown moderate to strong interclass correlation (r=0.73, P

___

  • 1. Boskey A, Vigorita V, Sencer O, Stuchin S, Lane J. Chemical, microscopic, and ultrastructural characterization of the mineral deposits in tumoral calcinosis. Clinical orthopaedics and related research 1983: 258-269.
  • 2. Salomon LJ, Alfirevic Z, Berghella V, Bilardo C, Hernandez‐ Andrade E, Johnsen S, et al. Practice guidelines for performance of the routine mid‐ trimester fetal ultrasound scan. Ultrasound in Obstetrics & Gynecology 2011; 37: 116-126.
  • 3. Obstetrics ISoUi, Committee GE. Sonographic examination of the fetal central nervous system: guidelines for performing the'basic examination'and the'fetal neurosonogram'. Ultrasound in obstetrics & gynecology: the official journal of the International Society of Ultrasound in Obstetrics and Gynecology 2007; 29: 109-116.
  • 4. Guibaud L. Fetal cerebral ventricular measurement and ventriculomegaly: time for procedure standardization. Wiley Online Library; 2009.
  • 5. Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. Journal of biopharmaceutical statistics 2007; 17: 571-582.
  • 6. Tilea B, Alberti C, Adamsbaum C, Armoogum P, Oury J, Cabrol D, et al. Cerebral biometry in fetal magnetic resonance imaging: new reference data. Ultrasound in Obstetrics and Gynecology 2009; 33: 173-181.
  • 7. Garel C. Fetal cerebral biometry: normal parenchymal findings and ventricular size. European radiology 2005; 15: 809-813.
Eastern Journal of Medicine-Cover
  • ISSN: 1301-0883
  • Başlangıç: 1996
  • Yayıncı: ERBİL KARAMAN